Contact US For Quotation | Email:mewoi009@mewoi.com | Whatsapp&Wechat:+8619925688679 |
Application
OTR-1903 is based on the coulometric sensor method and available for the OTR test of flexible films, sheeting and packages. |
Features
Coulometric sensor method Own technology Film and package test one piece specimen test One-button test, test automatically in whole process Multiple test modes optional Data curves display Computer control Digital flow adjustment Chamber with temperature sensor embedded in Environment temperature and humidity sensor inside Chambers with water bath temperature control technology Unit based on embedded system,24bit Δ-Σ AD Pressure, oxygen and temperature overload protection Traceable reference film calibration Supports DSM system (DSM, lab data management system) |
Technical
Test range | Film: 0.01-6300 cm3/m2·d (Standard) 0.07-6,3000 cm3/m2·d (Optional) Resolution: 0.001 cm3/m2·d Package:0.0001~62 cm3/pkg·d Resolution: 0.00001 cm3/m2·d |
Specimen amount | one piece |
Temperature range | 5C – 95C (with temperature controller) |
Temperature accuracy | ±0.1C |
Humidity range | 0%RH,35%RH-90%RH,100%RH |
Humidity accuracy | ±1%RH |
Test area | 50cm2 |
Specimen size | Film: size:>=150mm * 94mm; thickness: less 1mm Package: max size: D120mm * 400mm |
Carrier gas | 99.999% Nitrogen |
Carrier flow | 0~200ml/min |
Gas interface | 1/8inch |
Power supply | AC 220V 50Hz |
Dimension | 330mm(L) * 600mm(W) * 330mm(H) |
Net weight | 28kg |
Standard
ASTM D3985, ASTM F2622, ASTM F1307, ASTM F1927, ISO 15105-2, GB/T 19789, JIS K7126-B, YBB 00082003 |
Configuration
Delivery: Main unit, computer, OX-TRANS software, tank regulator, PCIE card, communication cable, sample cutter, vacuum grease. Optional: Temperature controller, package fixtures, reference film, vacuum grease, sample cutter, DSM system. Note: Specifications are subject to change without prior notice. |